In the menu of the HM8112-3 you can adjust the time between two measurements from 0.01s to 60s. This means that one measurement value will capture every 0.01s or 60s.

 

The measurement rate can be adjusted with following values:

 

- 10ms (over interface only)

- 50ms (over interface only)

- 100ms (Default after switching on the power)

- 500ms

- 1s

- 10s

- 60s

 

A display rate of 60s means:

The HM8112-3 integrates the input voltage and the the reference voltage over a period of 60s. After expiry of this time the value calculated will be displayed by 6½ digits.

 

More detailed information you can find in the manual of the HM8112-3.