The HAMEG RS-232 interface is arranged as a 9-pole D-SUB jack. This bidirectional interface is intended to send and receive setup parameters, data and remote commands from a PC (serial port) to the HAMEG device.

 

A connection to the PC can be established with a 9-pole, shielded (1:1 wired) cable with D-SUB plugs on each end.

 

The PIN assignment:

 




2 Tx Data (Data from the HAMEG device to the PC) 
3 Rx Data (Data from the PC to the HAMEG device) 
7 CTS Ready to send 
8 RTS Ready to receive 
5 Ground (Reference potential connected with HAMEG device (safety class I) and power cable to the grounding conductor) 
9 +5 V Supply voltage for external devices (max. 400 mA) 
  
  
HO820:  You don't have +5V on Pin 9 (Pin 9 unconnected)! 

 

 

The maximum voltage at the terminals Tx, Rx, RTS and CTS is ±12V.

 

The standard RS-232 paramteters are:

 

Oscilloscopes:

8-N-2 (8 Data bits, no parity bit, 2 stop bits)

 

Programmable Measuring Instruments:

8-N-1 (8 Data bits, no parity bit, 1 stop bit)

 

RTS/CTS Hardware protocol: None.

These parameters may be changed with the device.